We went through the Chapter 14 text line by line. There is no provision capping chamber
oxygen concentration at 23.5%. The number appears in three roles: the Chapter 3
definition of an oxygen-enriched atmosphere, an oxygen-monitor alarm setpoint at
14.2.9.4.2.1, and a trigger for antistatic procedures and patient grounding at
14.3.1.5.3.1–.2.
Read 14.3.1.5.3.2 closely and it is the opposite of a cap: it prescribes extra
precautions for Class A and Class B chambers running above 23.5%. Single-occupant
Class B chambers commonly run far above it by design. Asked directly about the figure, the
UHMS Hyperbaric Oxygen Safety Committee replied that in their opinion it
relates to the Class A chamber, not the room housing it.
Where 23.5% is a hard regulatory limit is OSHA's confined-space rule, which
classifies an atmosphere above 23.5% oxygen as
hazardous for worker-occupied spaces — a
different code, a different scope, and about the room your staff stands in rather than the
vessel.
We would rather lose the simpler sales line than repeat a claim we cannot cite.